LB 470 Level 9 Troubleshooting
56925BA2 Rev. 03, 11/2017
M107
Error in the internal real
time clock
M
Malfunction of the real-time clock.
Check Date and Time. If the event oc-
curs frequently, contact Berthold
Service.
M108 CPU temperature sensor M
The temperature sensor of the device is
defective. Contact Berthold Service. The
hardware is defective and, if necessary,
must be checked and replaced.
M109
Maintenance required
M
The internal temperature of the device
is close to the lower threshold value of
the permissible operating temperature.
M110 Temp LL OOS S
The internal temperature of the device
is below the lower limit. The correct
function of the device can not be guar-
anteed. It is recommend to have the
device checked by Berthold Service,
even if it seems to work normally.
M111 Temp UL maintenace M
The internal temperature of the device
is close to the upper limit.
M112 Temp UL OOS S
The internal temperature of the device
is above the upper limit. The correct
function of the device can not be guar-
anteed. It is recommend to have the
device checked by Berthold Service,
even if it seems to work normally.
M113 Power On Reboot C
The device was restarted due to un-
known reasons.
M114 Software Reboot C The device was restarted by user input.
M115 Extern RTC malfunction M
Failure of the external real time clock.
Contact Berthold service, if this event
occurs repeatedly.
M116 Corrupt Date M
The date could not be verified at
startup. Check date and time and set if
necessary.
M107 Intern RTC malfunction M
Failure of the real time clock. Verify
date and time. Contact Berthold service,
if this event occurs repeatedly.
M108 CPU temp sensor M
Failure of the device temperature sen-
sor. Contact Berthold service. The
Hardware is defective and needs to be
checked / replaced.
M109 Temp LL maintenance M
The internal temperature of the device
is close to the lower limit.