BestCode Next Series 8 Technical Manual October 2022 Page 165 of 290
Prompt Code
Prompt Name
Prompt Description
04-8001
Internal
Memory device not detected.
Displayed to indicate a fatal error with the internal file system (NOR).
Solution 1:
Install USB and Scan External device. Then Erase and Reformat Memory.
Solution 2:
Load Firmware
Solution 3:
Replace the main Circuit Board
Prompt Code
Prompt Name
Prompt Description
04-8002
FPGA Load Fault
The FPGA did not load properly.
Displayed to indicate a fatal error loading the FPGA.
Solution 1:
Load Firmware
Solution 2:
Replace the main Circuit Board
Prompt Code
Prompt Name
Prompt Description
04-8003
Memory Fault
"The message is out of memory."
Displayed to indicate that an element could not be allocated to add to a message being created or
editted.
Solution 1:
Message is at maximum length. Adjust message to reduce number of fields then try again.
Prompt Code
Prompt Name
Prompt Description
04-8004
Script
"Process script failed: %ls"
Displayed to indicate that a script could not be parsed.
Solution 1:
Install USB and Scan External device. Then Erase and Reformat Memory.
Solution 2:
Reload Firmware.
Prompt Code
Prompt Name
Prompt Description
04-8005
NOT USED
Solution 1:
Update Firmware Version
Prompt Code
Prompt Name
Prompt Description
04-8006
File Transfer
Translator file format error: %ls
Written to the event log to indicate that the format of a translation file was not valid.
Solution 1:
Scan external device to ensure USB is not corrupted. Perform Restore function
(Service>Tools>Restore)
Solution 2:
Translation files are not saved correctly. Re-create and save the translation file correctly.
Prompt Code
Prompt Name
Prompt Description
04-8007
Process Failed
Import script failed: %ls
Displayed to indicate that the specified script could not be read.
Solution 1:
Scan external device to ensure USB is not corrupted. Perform Restore function
(Service>Tools>Restore)
Solution 2:
Load Firmware Via the J-Link tool
Prompt Code
Prompt Name
Prompt Description
04-8008
Charge
"Print file not found: %ls"
Indicates that the specified file could not be found.
Solution 1:
Scan external device to ensure USB is not corrupted. Perform Erase and Reformat Memory.