EasyManua.ls Logo

Binder KB - Temperature Unit Selection; Data Recording Rate

Binder KB
147 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
KB / KB-UL (E4 + E6) 10/2021 page 86/147
12.5 Changing the temperature unit
To select the temperature unit, go to Menu > Settings > Temperature unit
Fixed value
08.03.2015 05:05:06
Submenu “Temperature unit”.
Select the desired temperature unit”
and press the operating button.
..\ Temperature unit
Celsius [°C]
Fahrenheit [°F]
Kelvin [K]
Close Home
If you selected the temperature unit, the controller returns to the „Settings“ menu..
Otherwise, go back to the “Settings” menu with “Close” or to the initial view with “Home”.
12.6 Defining the data recording rate
To define the recording rate for data storage on the SD card, go to Menu > Settings > Recording rate
Fixed value
08.03.2015 05:05:06
Entry menu “Recording rate”.
The current recording rate is shown. Enter the de-
sired value with the operating button.
Shortest possible value: 60 seconds.
Press the “Ok” button to confirm.
..\ Recording rate (seconds)
6
0
0
Ok 0 1 2 3 4 5 7 8 9 . Del Ins
Close Ok Home
When entering a value smaller than 60 seconds, the message “invalid value” appears. Press the operating
button to confirm with “Ok” and repeat the entry with a correct value.
Go back to the “Settings” menu with “Close” or to the initial view with “Home”.
Note:
After a period of 9 years, the controller starts overwriting the oldest values on the storage medium. This is
independent of the selected storage interval and the actual operating time of the chamber. In any case, the
data can be read out at any time using the function “Export to USB drive” (chap. 13.1) and stored externally.