17
IMPORTANT: If a value increase or decrease is not possible due to a hard limit (as 19.5
could be expected in this case), the value of the hard limit is displayed as the highest
acceptable value.
If the down arrow
d is pressed on the above screen, the pump displays the previous
value i.e. 9.5.
ENTERING VALUES
If the up arrow u is pressed on the
above screen, the pump displays a clear
message that the value entered exceeds
the hard limit.
After the user confirms the message by
pressing the OK button k, the pump pre-
fills the programming editor field with the
last value programmed and confirmed
prior to the value that created the alert.
While in the editor function the arrow
buttons Q and the OK button k are
used to define and confirm a value, the C
(clear) button c can be used to reset the
value to 0.
To exit the editor function, press the C button c again. The value of the infusion parameter
reverts to the last confirmed value.