Note: During priming, all air in line alarms are inactive.
•
Connect IV line to the patient and observe B. Braun landing page. Press k to
program infusion.
Note: Pump automatically powers up in the drug library for all new infusions.
Note: Last infusion, if any, will be displayed with prompt "Continue last infusion?".
Press u for yes or dfor no. If yes, confirm each infusion parameter by scrolling with
t and pressing k to confirm values. If no, press kon B. Braun landing page to
program new infusion.
1.2 Drug Library
The Space Infusomat powers up in the drug library, providing a safety net for clinicians
during programming.
The drug library allows for up to 1200 drug names in up to 10 concentrations per drug
to be sub-divided into 30 categories. These drugs can be assigned to 50 care units and
16 Pat
ient Profiles. Drug categories allow the drug names to be subdivided, such as by
drug type. Patient profiles allow one drug to have different settings such as dose limits,
concentration, clinical advisories, pressure limits, and infusion type based on patient
type or condition. Care Units allow drugs to have different settings based on location
or classes of patients.
•
The pump will prompt user to enter weight for weight based dosing and BSA for m
2
dosing.
Note: Changing the weight during a weight based infusion or BSA during a m
2
infusion will result in a change in the infusion rate to deliver the programmed dose at
the new weight.