Cleaning and Disinfecting
Caution: Before cleaning and disinfecting the pump, always disconnect
the pump from the patient, switch off the device and disconnect pump from A/C
power outlet and other devices (e.g. staff call).
Clean all external surfaces using a clean, soft, lint-free cloth dampened with a
mild cleaning solution of soapy water. Make sure to remove any visible residue
from all surfaces pri
or to
disinfecting. The housing of Infusomat Space , the line
guide cover and the IV line guiding areas may be disinfected with EPA registered
hospital disinfectants containing 1-propanol, isopropyl alcohol, ethanol or
didecyl dimethyl ammonium chloride. Do not spray disinfectants directly on the
pump, use a soft, low lint cloth dampened but not saturated with product. After
cleaning and disinfecting a
llo
w device to dry for at least 20 minutes prior to
use.
Note: To clean and disinfect the peristaltic fingers and membrane, the line guide
cover can be removed using a pointed object (ballpoint pen) inserted in the
lower right corner. Clean cover and peristaltic fingers and membrane using a
clean, soft, lint-free cloth dampened with a mild cleaning solution of soapy
water. The line guide cover may be cl
ean
ed under running water after it is
removed from the pump. Disinfect line guide cover, peristaltic fingers and
membrame following same procedure as stated above.
Caution: Do not touch the peristaltic fingers and membrane area of
pump with sharp object.
When reinserting the line guide cover, make sure that it is not damaged and
that it audibly locks in place.
Note: Do not use chloride disinfectant pro
ducts (b
leach).
line guide cover
Chapter 9
WARRANTY / TRAINING / TSC* / SERVICE / DISINFECTING / DISPOSAL

Table of Contents

Other manuals for Braun Infusomat Space

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Infusomat Space and is the answer not in the manual?

Braun Infusomat Space Specifications

General IconGeneral
TypeInfusion Pump
Power SupplyAC/DC adapter, rechargeable battery
DisplayColor graphic display
AlarmsAudible and visual alarms
Flow Rate Accuracy±5%
Occlusion PressureAdjustable
VTBI Range0.1 – 9999 ml
KVO Rate0.1 – 5 ml/h (adjustable)
BatteryLithium-ion
Operating ModesContinuous
CertificationsCE, ISO 13485, IEC 60601-1