48
• Note that upon activation of data lock the y symbol appears on the run
screen to the right of the rate/dose indicating changes are only possible after
entering the code. If the wrong code is entered four times the pump will go
into an audible alarm, the yellow LED will light, and the pump display
indicates invalid code.
• To deactivate data lock, select “Off” in the data lock menu, press k.
2.2.6 KVO-Mode
The pump can continue the infusion with a preset KVO rate after an infusion time or
VTBI has ended. The rate and duration of the KVO delivery is set in the service program.
When KVO feature is activated in the service program, the pump will automatically go
into KVO unless it has been turned off in the Options menu.
•
Open the KVO mode in the Options menu with l.
•
Answer the Yes/No question with u to activate the KVO mode.
Note: KVO function is disabled in Dose over Time.
2.2.7 Contrast / Display Light / Keypad Light
Contrast as well as display and keypad light can be adjusted individually according to
the lighting conditions.
•
Open contrast/display light/keypad light in Options Menu by pressing l.
•
Choose between 9 contrast and display light levels with l or r and confirm
with k.
Chapter 2
PUMP MENUS