The Infusomat® Space is equipped with an audible and optical alarm signal.
3.1 Device Alarms
When a device alarm occurs the infusion is immediately stopped and display indicates
"device alarm" with a code. The audible alarm is unique. Press o to switch off the
device. Then switch the device on again. In the case of a repeated device alarm the
pump must be sent for service.
3.2 Pre-Alarms and Operating Alarms
Pre-alarms:
Pre-alarms are set in the service program. Pre-alarms occur a few minutes (specific
time set in service settings) prior to operating alarms. During pre-alarms an audible
tone sounds, the yellow LED is constantly on and a staff call is activated (optional). The
display message varies depending on the reason for the alarm. The signal tone and the
staff call are turned off with
k. Display and LED stay in pre-alarm until condition
causing the pre-alarm results in an operating alarm. Pre-alarms do not stop the infusion.
Display message Pre-alarm reason
“VTBI near end“ The programmed volume is almost infused.
“Time near end“ The programmed time is almost over.
“Battery nearly empty“ The battery is almost discharged.
“KVO mode“ VTBI or time are complete and the pump
converted to the KVO rate.
Alarm-
type
Device
Alarm
yes flashes off device alarm
and alarm
code
yes Follow the instruc-
tion on the display
to press k.
Opera-
ting
Alarm
yes flashes off alarm type yes
Pre-
Alarm
yes off constant
on
alarm type (de-)activate
via service
program
Press k to acknow-
ledge and cancel
alarm
Reminder
Alarm
yes off constant
on
alarm type yes
Press k to acknow-
ledge and cancel
alarm
Alarm
Hint
no off off alarm type no
Message disappears
without confirmation.
Audible
signal
Red LED
Optical signal
Yellow LED Text
Staff call User confirmation
Press k to
acknowledge and
cancel alarm