EasyManua.ls Logo

Braun Infusomat Space - Technical Data; Chapter 8 Technical Data

Braun Infusomat Space
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
67
TECHNICAL DATA
Type of unit Volumetric infusion pump
Classification (acc. to IEC/EN 60601-1) defibrillator-proof; CF equipment
Protective Class II; Protective Class I in
combination with SpaceStation
Class (acc. to Directive 93/42 EEC) IIb
Classification acc. to 21 CFR 880.5725 Class II (Product code FRN and FPA)
Moisture protection IP 22 (drip protected for horizontal usage)
External power supply:
Rated voltage Via B. Braun SpaceStation or optional
AC adaptor (rated voltage 100 240 V
AC~, 50/60 Hz) for stand alone operation
External low voltage 11 16 V DC via Connection Lead SP
12 V or via SpaceStation
Staff call Max. 24 V / 0,5 A / 24 VA
EMC IEC/EN 60601-1-2 / 60601-2-24
Time of operation 100 % (continuous operation)
Operating conditions:
Relative humidity 30 % 90 % (without condensation)
Temperature +60 + 105° F
Atmospheric pressure 500 1060 mbar
Storage conditions:
Relative humidity 20 % 90 % (without condensation)
Temperature -4 +131° F
Atmospheric pressure 500 1060 mbar
Type of battery pack (rechargeable) Li-Ion
NiMH
Operating time of rechargeable battery
Li-Ion wireless active Infusomat at 100ml/h
4.0 hours
wireless active Infusomat at 1200ml/h
2.5 hours
wireless inactive Infusomat at 100ml/h
12.0 hours
wireless inactive Infusomat at 1200ml/h
5.0 hours
NiMH at 100 ml/h 13 hours
at 1200 ml/h 5 hours
Recharging time Approximately 6 hours
Weight Approximately 1.4 kg
Dimensions (W x H x D) 214 x 68 x 124 mm = 8.4 x 2.6 x 4.8 inches