67
TECHNICAL DATA
Type of unit Volumetric infusion pump
Classification (acc. to IEC/EN 60601-1) defibrillator-proof; CF equipment
Protective Class II; Protective Class I in
combination with SpaceStation
Class (acc. to Directive 93/42 EEC) IIb
Classification acc. to 21 CFR 880.5725 Class II (Product code FRN and FPA)
Moisture protection IP 22 (drip protected for horizontal usage)
External power supply:
• Rated voltage Via B. Braun SpaceStation or optional
AC adaptor (rated voltage 100 … 240 V
AC~, 50/60 Hz) for stand alone operation
• External low voltage 11 – 16 V DC via Connection Lead SP
12 V or via SpaceStation
Staff call Max. 24 V / 0,5 A / 24 VA
EMC IEC/EN 60601-1-2 / 60601-2-24
Time of operation 100 % (continuous operation)
Operating conditions:
• Relative humidity 30 % – 90 % (without condensation)
• Temperature +60 – + 105° F
• Atmospheric pressure 500 – 1060 mbar
Storage conditions:
• Relative humidity 20 % – 90 % (without condensation)
• Temperature -4 – +131° F
• Atmospheric pressure 500 – 1060 mbar
Type of battery pack (rechargeable) Li-Ion
NiMH
Operating time of rechargeable battery
Li-Ion wireless active Infusomat at 100ml/h
4.0 hours
wireless active Infusomat at 1200ml/h
2.5 hours
wireless inactive Infusomat at 100ml/h
12.0 hours
wireless inactive Infusomat at 1200ml/h
5.0 hours
NiMH at 100 ml/h 13 hours
at 1200 ml/h 5 hours
Recharging time Approximately 6 hours
Weight Approximately 1.4 kg
Dimensions (W x H x D) 214 x 68 x 124 mm = 8.4 x 2.6 x 4.8 inches
TECHNICAL DATA
Chapter 8