33
• Enter Total time and press k to confirm.
Note: When default values have been set in the Drug Library there will not be a prompt
to enter values. Values may be edited by scrolling to the parameter and pressing the
l arrow key.
Note: The KVO and Bolus functions are disabled during Dose Over Time.
1.6 Programming a Loading Dose in Drug Library
The Infusomat® Space allows for the delivery of a loading dose for medications that
have been set up in the drug library for loading dose. Loading doses may have soft and
hard limits set in the Drug Library. The pump will prompt and allow the user to pro-
ceed with a loading dose or go right to the continuous infusion. Once "No" has been
selected to "Program Loading Dose?" it is not possible to recall it on the pump without
exiting and clearing the infusion by
pressing the c key and re-programming the
drug from the beginning.
To Program a loading dose:
•
Press u arrow to answer yes to loading dose prompt.
• Enter loading dose amount using q arrow keys, confirm with k.
• Enter loading dose time using q arrow keys, confirm with k.
Chapter 1
OPERATION