Note: Soft and Hard limit alerts will be generated based on drug library settings.
Press l arrow key to access doserate/rate editor to set continuous infusion.
Enter doserate or rate for continuous infusion and confirm with k.
Enter VTBI or Time.
Note: "LD dose and Start" alternate in upper right corner of screen.
Press sf to begin infusion.
Note: The word LOAD is super-imposed on the run arrow
s.
Note: The loading dose may be stopped at any time by pressing the sf key. Press sf
key to deliver remaining loading dose. There will be nothing infusing until Load Dose is
resumed or cancelled and continuous infusion started.
When pump is stopped intentionally or by an alarm state, the pump will
display the amount that has been delivered of the total amount
programmed. Press k to confirm.
Upo
n p
ressing sf , the pump will prompt to deliver remaining loading dose.
Press u arrow for yes, d arrow for no.