40
Chapter 1
OPERATION
3.) Enter rate and time: The infusion volume will be calculated and displayed on the
bottom of the display.
Rate titration will result in re-calculation of VTBI.
1.12 IV Line Change and New Therapy Start
Note: Always close roller clamp and disconnect the line from the patient before
changing a line to avoid inadvertent free flow.
Press sf to stop the delivery. The green LED goes out. Close the roller clamp
and disconnect line from the patient.
Press x and open the pump door with u. Once door finishes opening
automatically, pull down on door to fully open. Expect some resistance. Press down
the green opening lever completely until it locks in place, releasing the pump based
safety free flow clamp, and yellow light flashes, then remove the IV line from right
to left and insert a new IV line,
note that green anti-free flow clip occludes the
IV line when removed.
Note: In the unlikely event the pump door cannot be opened remove the allen key
from the inside of the battery compartment cover. Use this key to remove the
emergency aperture cover of the pump. Place the crank in the aperture and turn it
clockwise until the pump door opens.
Push cover opening with pen.
Remove crank from inside of battery
cover.