Note: The word Bolus superimposed over the run arrows is displayed on the screen.
Note: Failure to press BOL to start bolus will result in reminder alarm.
• The pump will automatically convert back to the continuous infusion when
the bolus is complete.
Note: The bolus may be stopped at any time by pressing the c key, the
continuous will then run. Press nb key to deliver remaining bolus.
• Press sf to sto
p the infusion entirely.
• When bolus is stopped intentionally by pressing c key,or pump is stopped
by an alarm state, the pump will display the amount that has been delivered
of the total amount programmed. Press k to confirm.
• If pump was not running, as after an operating alarm, press sf to begin the
infusionand respond to prompt to deliver remaining bolus per above.
The last bolus dose is r
ecorded in the status menu, refer to ch. 2.3.
1.8 Manual Bolus
The Infusomat® Space allows for the delivery of a manual bolus when the drug is set
up in the drug library for a manual bolus. A manual bolus requires the user to
continually hold the bolus button to deliver the bolus.
Note: Drug library limits do not apply to manual boluses.