Chapter 7
START UP GRAPHS AND TRUMPET CURVES
The graphs show the accuracy/uniformity of flow in relation to time.
The delivery behaviour or delivery precision is influenced by the type of the
disposable used. Deviations from the technical data of the pump cannot be
excluded if lines (disposables) other than B.Braun approved sets are used.
START UP GRAPHS AND TRUMPET CURVES
Trumpet Curves
Measured values for second hour in each case.
Measurement interval t = 0.5 min
Observation interval p x t [min]
Start Up Graphs
Measurement interval t = 0.5 min
Measurement duration T = 120 min
Flow Q
i
(ml/h)
Start Up Graphs Trumpet Curves