49
2.2.8 Bolus Rate
The pump has a default bolus rate which is set in the service program. This rate is used
for manual bolusing. For a programmed bolus, this rate will be converted to a time in
the time editor screen if no default bolus rate has been set in the drug library and may
be changed by adjusting the time.
•
Open bolus rate in Options Menu with l.
•
Change bolus rate with q and confirm setting with k.
2.2.9 Date / Time
• Open date/time in the Options Menu with l.
•
Modify date and time with q and confirm the setting with k.
2.2.10 Macro Mode
The infusion rate appears much larger and the drug name much smaller on the display
when the macro mode is activated and the pump is infusing.
•
Open macro mode in Options Menu with l.
•
Answer Yes/No question by pressing u to activate the macro mode.
Note: For quick activation and deactivation of macro mode: Press and hold r while
the pump is infusing until the font size changes.
2.2.11 Wireless Activation
Allows wireless to set for active or inactive.
2.3 Status Menu
The status menu is accessed from the Home screen. In the status menu it is pos-
sible to review the following:
•
Battery time remaining at current infusion rate
•
Last bolus amount, date and time
•
Drug info which includes Care Unit, drug file creation date, current drug selection,
patient profile and clinical advisory (if any).
•
Pump software version
•
Wireless status
•
IV line
Chapter 2
PUMP MENUS