To deliver a manual bolus:
While an infusion is running, press nb key.
Select “Use Manual Bolus feature”.
Press and hold nb key.
Bolus amount will count up on pump display.
Manual bolus is limited to 10 seconds.
The pump may be set to emit an audible tone every 1ml of solution delivered.
Bolus amount will be displayed when 10 seconds is complete or nb key is released.
The pump converts to the continu
ous infusion when the nb key is released or
10 seconds is completed.
1.9 Changing Care Unit
The Infusomat® Space allows the Care Unit to be changed when patients are transfer-
red. All drug library limits for the new Care Unit are immediately applied and a limit
alert appears if soft limits are exceeded in the new Care Unit. If a hard limit is exceeded
the pump reverts back to the previous Care Unit.
The pump does not allow a Care Unit
change in the following circumstances: during SECondary infusions, if the drug or
concentration are not available in the new Care Unit and if hard limit is exceeded.
To change a Care Unit:
While infusion is running press the c key to get to home screen.
Scroll to “Change Care Unit”, select with l arrow or k key.
Scroll to desired Care Unit and select with left l o
r k key.
Pump will display confirmation of new Care Unit.