Chapter 1
• Press x and open the pump door with u and remove IV line per above in
Section 1.12.
•
Close door and press o for 3 sec to power off the pump.
Note: When pump is powered up user will be prompted to continue last infusion,
answering No returns to B. Braun landing page.
Note: Pump cannot be powered off with IV line inserted.
1.14 Standby Mode
The pump may be placed in standby rather than powering off so that re-starting an
infusion is quicker.
•
Press sf to stop the infusion, leave IV line inserted in pump. Then press and
hold o for 3 sec.
•
The pump is now in Standby.
While the pump is in the standby mode, the display shows infusion mode, drug name
and the remaining time for standby mode. Change remaining time by pressing l,
standby may be set from 1 min to 24 hours. Exit standby by pressing c. The pump
wi
ll alarm when the standby time expires.
•
Press c to cancel standby.