OPERATION
The doserate, dose or time editor will appear and required confirmation if parameter
exceeds the soft limit in the new Care Unit. The soft limit warning will be shown
requiring an over ride or new programming to proceed as in Section 1.2. Pump will
display message if Care Unit was not changed because drug or drug concentration is
not available in new Care Unit or hard limits are exceeded.
1.10 Changing display while pump is running
While pump is running, press d arrow to choose preferred value to display
on bottom left corner of display.
Values displayed will vary with type of therapy and may include drug long name, con-
centration, volume totals, remaining time, care area, etc. Displayed value will remain
until changed by user.
1.11 Basic Infusion (programming the pump outside of
the d
rug library)
The Infusomat® Space allows programming outside the drug library, referred to as a
Basic infusion. To program a Basic infusion, select Basic Infusion” from the Care Unit,
Patient Profile, Drug Category or Drug selection screens.
Note: Basic infusion may be selected in the SECondary menu.
The pump provides a prompt that programming outside the drug library has
no safety limits, pres