Chapter 4
WIRELESS DRUG LIBRARY UPLOAD
WIRELESS DRUG LI BRARY UPLOAD
The pump has the ability to accept new drug library files wirelessly. A file
symbol will flash alternately with the wireless antenna symbol on the top of the
pump display when a new file is available. The wireless antenna symbol is seen
on run screen, standby screen and when pump is powered off and plugged in.
Press Start/Stop key sf to stop infusion when patient condition allows.
Close roller clamp, disconnect IV line from patient, press x and remove
IV line from right to left per Section 1.12 and 1.13.
Power pump off.
Wait 10 seconds, progress bar appears on pump, do not power pump back
on until Drug Upload is complete as indicated by progress bar.
Display will indicate DL update is successful when complete, on power up
display indicates new Drug Library is activated on pump.
Note: Cancelling the Drug library update will remove all drug library files from
the pump. A small drug library may load very quickly and not be able to be
cancelled.