55
Chapter 4
WIRELESS DRUG LIBRARY UPLOAD
WIRELESS DRUG LI BRARY UPLOAD
The pump has the ability to accept new drug library files wirelessly. A file
symbol will flash alternately with the wireless antenna symbol on the top of the
pump display when a new file is available. The wireless antenna symbol is seen
on run screen, standby screen and when pump is powered off and plugged in.
•
Press Start/Stop key sf to stop infusion when patient condition allows.
•
Close roller clamp, disconnect IV line from patient, press x and remove
IV line from right to left per Section 1.12 and 1.13.
•
Power pump off.
•
Wait 10 seconds, progress bar appears on pump, do not power pump back
on until Drug Upload is complete as indicated by progress bar.
•
Display will indicate DL update is successful when complete, on power up
display indicates new Drug Library is activated on pump.
Note: Cancelling the Drug library update will remove all drug library files from
the pump. A small drug library may load very quickly and not be able to be
cancelled.