18
OPERATION
1.1 Inserting IV Line and Priming in Pump
• Ensure that the pump is properly installed (see Infusomat Space Overview above).
Check the equipment for completeness and damages.
•
Spike infusion bag and fill drip chamber to 2/3 full.
•
Press oto power on. The message “Self-test active” and the software version are
displayed, two audible tones sound and all three LEDs (yellow, green/red and blue)
flash once. The power supply indicator and the set pressure level are
displayed followed by the set limit for the accumulated air volume and the max.
size of air bubbles.
•
Observe B. Braun landing page and press k to enter infusion parameters followed
by IV line insertion, or press x and then press u to open the pump door and
insert the line.
Note: Observe message indicating the roller clamp should be closed during set change.
C
aution: Close the roller clamp before inserting the IV line and do not
connect to patient until properly loaded and primed.
Chapter 1
OPERATION
Insert the IV line from right to
the left starting with the 2 hole
clip.