19
• Firmly press tubing into the air sensor guide to make sure the line is properly
inserted into the sensors. Thread the tubing through the notches on the right and
left side of the pump.
•
Close the pump door by firmly placing pressure with both hands on each side of the
pump door, continue to press firmly until you hear and feel the motorized door
latch pull the door shut. Do not open roller cl
amp until the pump directs you to do
so when self test is completed.
Caution: Pump door will not close if the anti free flow clip is not
properly inserted. Do not force door closed - if door is too difficult to
close, check IV line and anti free-flow clip (green) for proper loading.
Caution: Before opening door, close roller clamp. If the door hook
appears damaged or broken remove the pump from s
ervice
.
•
If priming is activated, press u to select yes to prime IV line and note prompt to
disconnect IV line from patient before priming. Press k to confirm and begin
priming. Respond to "priming stopped" message by confirming with k. Respond to
"repeat priming" prompt with u for yes if line is not fully primed or dfor no if
fully primed. Please refer to the priming volume on each disposable
label.
Chapter 1
OPERATION
Next attach the white clip.
Insure silicone segment is not
stretched or twisted, stars on tubing
must be in straight line and line
loading guide should not be twisted.
Insert the green anti-freeflow clip in
the green slot, in the direction indi-
cated by the arrow, until the opening
lever snaps out, yellow light goes out
and pump based free flow protection
safety clamp occludes the lines.
Anti-Free
Flow
clip insertion
Door hook
Air sensor
guide
Safety
clamp