2.2 Options
2.2.1 Downstream Occlusion Pressure
The higher the pressure level is set, the higher the pressure level must rise before trig-
gering an occlusion pressure alarm.
•
Enter "pressure" in Options Menu by pressing l to set downstream pressure limit.
•
Choose between nine pressure levels (1=lowest level; 9=highest level) by pressing
lor r and confirm entry with k. Pressure levels and equivalent mmHg are
displayed when left arrow is pressed while in pressure menu.
Note: The pressure will remain at set level until changed by user unless the drug
selected had a pressure level set in the drug library. When pump is powered off
pressure level returns to default value set in service program when powered back on
unless drug selected has a different pressure level set in the drug library.
2.2.2 Upstream Occlusion Pressure
The device is equipped with an upstream pressure sensor that detects an
occlusion (e.g. closed roller clamp, kinked line) between the container and the
pump. The higher the pressure level is set, the lower the pressure level must
drop before triggering an upstream occlusion pressure alarm.
• Access upstream pressure in Options Menu by pressing l.
• Choose between nine pressure l
eve
ls (1=lowest level; 9=highest level) by
pressing l or r and confirm entry with k.
2.2.3 Alarm Volume
Chose between 9 different alarm volume levels.
•
Open alarm volume in Options Menu with l.
•
Set volume with l or r and confirm entry with k.
2.2.4 Dose Rate Calculator
The Dose Rate Calculator may be used to calculate a doserate for a medication
that is not in the drug library. While the pump will calcu
late the rate, it is