47
Chapter 2
PUMP MENUS
2.2.5 Data Lock
The pump offers 2 levels of security to prevent unauthorized access which may
be set in this menu. A third level may be set in the drug library by drug. A four
digit code (default setting “9119”) must be entered within 20 seconds to prevent
a data lock alarm. The code can be changed via the service program for Level 1
and Level 2.
Level 1:
All keys except X door open and sf are locked and r
equir
e entry of data lock
code. The IV line may be changed.
Level 2:
Functions the same as level 1 and in addition does not allow the door to be
opened and requires code to start infusion.
Note: Once code has been entered changes may be made for 20 seconds until
the pump locks again and requires re-entry of the code.
Level 3:
Functions the same as level 2 but has a custom code set in the Drug Library. In
a
dditio
n the pump display may have a custom message.
Activation of the function:
• Open data lock in Options Menu with l.
• Select between level 1 or 2 with l and r and confirm with k.
• Enter code with q and press k in order to activate data lock.
Event
Change IV line
Level 1 Level 2 Level 3
Change parameters
Stop infusion
%
Powering off pump / Standby
%
Displays customized message when
running
= possible | = requires code | %= followed by data lock alarm
NA
NA
Start infusion
with code for
level 2/3