46
Select patient parameter, if any, for dosing calculation. Choices are weight,
BSA, or none.
Select the doserate units.
Enter doserate using q arrow keys and confirm with k.
Confirm and start infusion.
*Enter VTBI or Time.
To exit Dose Rate Calculation: the pump must be stopped. Press the c Key
from Home screen and answer “yes” to “Exit and clear infusion?”
Assigning a basic or doserate calcu
lat
ion infusion to Drug Library:
An infusion started without using the drug library, either a basic or doserate
calculation, may be assigned to the drug library without stopping the infusion.
Access the Home screen by pressing the c key.
Scroll to and select Options.
Scroll to and select “Assign to Drug Library”.
Program for the Drug Library following the same steps covered in Chapter
1.3 to p
rogram within the drug library, beginning with selecting the Care Unit.