52
A stopwatch on the display counts down the remaining time (depending on the
service program, between 3-30 min). After that, the pump goes into an opera-
ting alarm.
Operating alarms:
Operating alarms immediately stop the infusion. An audible tone sounds, the red
LED flashes and a staff call is activated (optional). The display states “Alarm“
and the reason for the operating alarm. The alarm tone and mes
sag
e as well as
the staff call are turned off with K. Correcting the alarm state depends on the
cause of the alarm.
Display message Alarm reason
“VTBI infused “ The programmed volume was infused.
Hang new bag and/or reset VTBI
“Time expired“ The programmed time has ended.
Hang new bag and/or reset time.
“Battery empty“ The battery is discharged.
The battery alarm will be on for 3 min.
Then the pump will automatically turn off. Plug
pump in immediately to re-charge battery.
"Downstream occlusion" The set downstream pressure level was exceeded. Post
occlusion bolus reduction is automatically initiated by
the pump.
Check tubing for kinks , closed stopcocks, filter
patency, and IV site.
Increase occlusion pressure if necessary per your
institutional policy.
“KVO time finished“ The KVO time has ended. Program new settings.
“Battery cover removed“ The battery cover is not properly engaged on the
battery compartment.
Reposition cover, listening for click when battery
cover is locked in place.
“Standby time expired“ The set standby time has ended.
Set new standby time or initiate infusion.
“No battery inserted“ It is not possible to use the pump without a
battery.
Turn off pump and insert battery according
to directions in “Overview Infusomat® Space“.
“Drive blocked“ Excess pressure in system or motor failure. Remove
tubing and re-insert, if persists pump must be sent for
service.
“Calibrate device“ Return to service for calibration.
Chapter 3
ALARMS