78
Note: Keep instrument upright and do not allow any part of instrument to
become saturated with or submersed in fluid during cleaning operation.
Do not allow moisture or detergents to come into contact with the electrical
connections of the device (P2 or P3 connectors) or any device openings. To
reduce the likelihood of moisture ingress into the electrical connectors, the P2
connector of a power suppl
y o
r combi cable may be used to cover the
connections during cleaning operations. Ensure that any connectors used to
cover are not connected to a wall outlet or other electrical source. Once the
cleaning has been completed, remove the connector and inspect all connectors
for residual moisture and evidence of damage or breakdown to the plating on
the connectors. Allow any residual moisture to evaporate
b
efore plugging the
device into a wall outlet. Replace any connectors which exhibit damage or evi-
dence of plating breakdown prior to returning the device to
service. Utilize electrical contact cleaner that does not react with plastics to
remove any deposits of material which may be present inside the electrical
connectors as required.
Caution: Do not allow liquids to enter into or come into contact
with
any openings or electrical connections on the pump or power supply.
Fluid exposure in these areas may result in the risk of short circuit,
corrosion or breakdown of sensitive electrical components, and/or
electrical shock. If fluid exposure occurs, the device should be swapped
out with another device in a manner that presents minimal interruption
to patient care. The device should rema
in
unplugged until it can be
inspected by a trained technician for any evidence of damage and/or
residual moisture which may impair the function of the device.
Caution: Do not touch line guide cover or peristalitic pumping area of
pump with sharp object.
When reinserting the line guide cover, make sure that it is not damaged and
that it audibly locks in place.
Note: The use of unappoved cleaners a
nd f
ailure to follow the disinfection pro-
cedures and the manufacturer's recommended dilutions can result in an instru-
ment malfunction or product damage and could void the warranty.
peristaltic fingers and membrane
Chapter 9
WARRANTY / TRAINING / TSC* / SERVICE / DISINFECTING / DISPOSAL