Example 3:
Pump maximum rate is 1200ml/hr
Current value 1100ml/hr
Pressing the uonce with the cursor in the hundreds column displays value
of 1200ml/hr which is at but not above the hard limit.
When the u is pressed once again with the cursor in the hundreds column
the hard limit message is displayed and remains until confirmed by pressing k.
Upon pressing k the pump editor field reverts to the last confirmed value
programmed prior to the value that created the alert.
Note: It is important to carefully check default values to be certain they match
physician order.
Chapter 1
OPERATION