26
Note: Care unit may be changed on any of the above navigation screens. This will re-
quire re-programming the therapy when a care unit change is done prior to beginning
the infusion. See Section 1.9 for instructions on changing the Care Unit while infusing.
Choose drug concentration using u and d arrows, press l arrow or k
to select.
Note: Some drugs may be set up for more than one infusion type, suc
h as Continuous
and Dose over Time, choose the mode using u and d arrows. Select with l or k
key. Refer to Section 1.5 for instructions on Dose over Time therapy.
After drug and concentration selection, the home screen will be displayed
with rate, doserate or total dose highlighted in white depending on the
infusion type and drug library settings. To program rate or doserate press
l arrow or k to op
en th
e editor screen. The total dose editor will be dis-
played for Dose over Time infusions.
When editing parameters, use the r and l arrow keys to move to the
digit to be programmed. The white background indicates current digit. The
c key may be used to clear existing values. Use u or d to program new
values. Text on the bottom of the screen indicates the parameter that
changes based on program
me
d value in the editor, as an example if rate is
being edited the time will appear on bottom of screen and will change based
on rate change (this requires VTBI to have been entered). The top of the
screen indicates infusion mode, confirm value with k , and soft limit
symbol.