29
"Back to PRIM" allows the pump to be set to require a manual switch back to PRIMary
in which case the pump stops and alarms when the SECondary is complete, or auto-
change whereby the pump converts to the PRIMary setting when the SECondary
dose/VTBI has been infused.
Note: The method of switching back to PRIMary remains for all SECondary infusions
until changed.
Select "New SECondary"
Select the drug
cate
gory and scroll to the desired drug name.
The Care Area and Patient Profile are retained from the PRIMary infusion. Select
infusion mode and concentration as done above whenprogramming a PRIMary
infusion.
Note: Medications may be set up in the drug library for SECondary only and will only
appear in the drug list when in the SECondary menu.
Note: PRIMary medications may be set in drug library to pro
hibit SECondary infusions.
A message will appear indicating SECondary has been disabled for the drug.
SECondary meds, which are administered intermittently, will most frequently be set up
for Dose over Time. SECondary meds set up as continuous are programmed as above
for PRIMary programming.
Enter total dose and confirm with k.
Enter total time and confirm with k.
Note: The pump calculates the rate
and VTBI.
Press sf key, note message reminder to check bag height and unclamp
secondary.
Press sf again to confirm message and start SECondary.
Chapter 1
OPERATION