31
Note: When running a SECondary infusion at rates greater than 300ml/hr clamp off
the PRIMary infusion to prevent concurrent flow.
Dose over Time run screen will indicate drug name and count down remaining time.
Note: During SECondary administration, the PRIMary may be resumed at any time by
pressing the sf key to stop the SECondary infusion, scrolling to and selecting "resume
PRIMary", selecting it wit
h l or k switches back to the primary infusion. Pressing
sf leads to an acknowledging prompt to clamp or remove SECondary, then pressing
sf key to infuse PRIM. To resume SECondary, stop PRIM and scroll to and select
"resume SECondary". The menu item "resume SECondary" will not appear if the
SECondary dose/volume was completed.
• To stop and clear a current SECondary infusion prior to completion of dose/
vo
lume, press sf key, then press c key, then press u arrow to answer
yes to "Clear SECinfusion". SECondary drug will remain in the list of previous
SECondary infusions.
The Space Infusomat stores up to 5 of the previously administered SECondary
medications for easier programming of repetitive SECondary medications. The
previous SECondary medications are cleared when the Care Unit changes, the pum
p is
powered off or answering “yes” to New Patient when bar coding.
Note: When a SECondary medication with the same name but different concentration
or dose is programmed, the prior medication with that name is deleted from the list.
SECondary infusions which are cleared prior to completion will remain in the list of
previous SECondary medications.
•
Stop PRIMary by pressing sf key.
• Scroll to SECondary
me
nu and select by pressing l arrow or k key.
• Select "Use a Previous SECondary" by pressing l arrow or k key.
Chapter 1
OPERATION