AUTOPROGRAMMING
Note: All normal pump functions remain in place when orders are received via
autoprogramming.
The pump can accept drug orders wirelessly from the EHR system. The workflow
to accept an order wirelessly will vary depending on your EHR vendor.
Using the hand held device or lap top, review the order and follow your hospital
protocol for scanning the bag, pump, patient and nurse (optional).
Once order is confirmed on the hand held or laptop, prompt EHR to send order
directly to pump. The order will arrive and appear on the pump within 10 seconds.
Ensure pump is on B Braun landing page (press ckey to return to landing page).
New Order message will appear with drug name and mode.
Press k to accept or ckey to cancel order and respond to prompt.
Select Care Unit and Patient Profile as in Drug Library programming in Chapter 1.
Pump will search for Drug Library match.
Note: If no drug library match, which may be due to no matching name, concentra-
tion or dosing units, pump displays reason for no match and depending on your
hospitals configuration either allows manual programming outside the drug library per
Chapter 1 Section 1.11 or rejects order completely. An order
that is confirmed outside
the drug library will have a triangle with an exclamation point on display to indicate
there are no drug library settings.
Scroll to each value to confirm using q arrow keys.
Note: Order may be cancelled prior to confirming order.
Chapter 5
AUTOPROGRAMMING AND BARCODING