59
Chapter 5
AUTOPROGRAMMING AND BARCODING
New Primary Infusion:
•
To accept a new PRIMary order, stop infusion and clear current PRIMary infusion by
pressing c key and responding “yes” to clear current infusion.
SECondary Orders:
Orders received after PRIMary has been set will be for SECondary infusions,
follow prompts on screen to stop the PRIMary to accept the SECondary order.
•
Confirm order values as above for PRIMary orders.
•
Respond to prompts to check bag height and clamps prior to starting SECondary.
New SECondary order while SECondary is Infusing:
•
Follow display prompts to stop current infusion.
Note: A SECondary order may be held for later by pressing c key to cancel
order and answering yes to “hold for later”.
Note: Changing values on any incoming order may only be done after confirm-
ing all values. Once all values are confirmed
yo
u may scroll to any value and
open editor with l to change value. Alternately, order may be cancelled and
request made for revised order to be sent.
Note: If pump is placed in standby while order is pending new order will flash
on top of stand by display, press M key to accept order (pump will come out of
standby).