59
Chapter 5
AUTOPROGRAMMING AND BARCODING
New Primary Infusion:
To accept a new PRIMary order, stop infusion and clear current PRIMary infusion by
pressing c key and respondingyesto clear current infusion.
SECondary Orders:
Orders received after PRIMary has been set will be for SECondary infusions,
follow prompts on screen to stop the PRIMary to accept the SECondary order.
Confirm order values as above for PRIMary orders.
Respond to prompts to check bag height and clamps prior to starting SECondary.
New SECondary order while SECondary is Infusing:
Follow display prompts to stop current infusion.
Note: A SECondary order may be held for later by pressing c key to cancel
order and answering yes to “hold for later”.
Note: Changing values on any incoming order may only be done after confirm-
ing all values. Once all values are confirmed
yo
u may scroll to any value and
open editor with l to change value. Alternately, order may be cancelled and
request made for revised order to be sent.
Note: If pump is placed in standby while order is pending new order will flash
on top of stand by display, press M key to accept order (pump will come out of
standby).