50
Chemyx, Inc.
Fusion Series User Manual
Returns: rate units, syringe inner diameter, transfer rate, priming
rate, time (in whole number minutes), transfer volume, time
delay (in whole number minutes)
status
Returns the current status of the pump. The returned integer
correlates to the status of the pump.
0: Pump stopped
1: Pump running
2: Pump paused
3: Pump delayed
4: Pump stalled
Returns: integer that represents the current status of the pump.
hexw2 X X X X X X start
Sets multiple parameters in one command (Basic Mode only).
X variables in order:
units: see set units command
transfer mode: 0-infusion, 1-withdrawal
syringe ID: see set diameter command
transfer volume: see set volume command; no negatives
transfer rate: see set rate command
start time delay: see set delay command
Example
The following example shows the
command using these parameters: units =
mL/min; transfer mode = withdrawal;
diameter = 23.05 mm; volume = 1.2 mL;
rate = 3.0 mL/min; delay = 30 sec (0.5
min)
hexw2 0 1 23.04 1.2 3.0 0.5
A start command can be added after the six parameters to
immediately start the run with the new parameters.
Example
hexw2 0 1 23.04 1.2 3.0 0.5
start
All six parameters do not need to be entered; however, all
parameters up to the parameter(s) of interest must be also be
included. The other parameters not entered will remain as they
are currently set.
Example