Fusion Series User Manual
set delay X,X,X,...
Sets the start time delay (in minutes) prior to each step of a
pump run. Upon starting the a step, the pump will wait until the
delay time elapses before starting to transfer volume.
X must be an integer or decimal (up to five decimal places)
within the rate limits for the syringe (See read limit parameter
command).
For Basic Mode, only one value is required. Entering only one
value will automatically switch the pump to Basic Mode.
Example
set delay 1.0
To set rates in Multi-Step Mode, the pump must be manually
switched to Multi-Step Mode on the pump. For each step of
Multi-Step Mode, a value must be entered, separated by
commas and no spaces.
Example
set delay 1.0,0.5,1.3
If the number of values is less than the number of steps, the
remaining steps will gain the same delay value as the last
entered value. Any extra values beyond the number of steps will
be truncated.
Returns the delay time (in Basic Mode).
Returns the delay time for each step (in Multi-Step Mode).
read limit parameter
Returns the min/max values that can be set for transfer volume
and transfer rate. The returned values are dependent on the
syringe inner diameter (ID). The transfer volume limits are
typically greater than the maximum volume of the syringe being
used.
Returns: max rate, min rate, max volume, min volume
dispensed volume
Returns the transferred volume for the current run/step. If no run
has started, returns the transferred volume for the last run/step.
elapsed time
Returns the elapsed time (in minutes) for the current run/step. If
no run has started, returns the elapsed time for the last run/step.
Elapsed time does not include any delay time.
view parameter
Returns the currently set parameters (Basic Mode only).