Chemyx, Inc.
Fusion Series User Manual
spaces. The number of steps will be automatically adjusted to
the number of values entered (i.e., entering only three values
will change the number of steps to three).
Example
set time 1.0,0.5,1.3
Returns target time and rate (in Basic Mode).
Returns no values (in Multi-Step Mode).
set rate X,X,X,...
Sets the transfer rate for infusion/withdrawal.
X must be an integer or decimal (up to five decimal places)
within the rate limits for the syringe (See read limit parameter
command).
For Basic Mode, only one value is required.
Example
set rate 1.0
To set rates in Multi-Step Mode, the pump must be manually
switched to Multi-Step Mode on the pump. For each step, a
value must be entered, separated by commas and no spaces.
Example
set rate 1.0,0.5,1.3
If the number of values is less than the number of steps, the
remaining steps will gain the same rate value as the last entered
value. Any extra values beyond the number of steps will be
truncated.
Variable rates may also be set in Multi-Step Mode. Rate 1 and
Rate 2 for each step can be set using a backslash. Each step is
still separated by a comma.
Example
set rate 0.5/1.0,0.5/1.0,1.3
Returns the flow rate and transfer time (in Basic Mode).
Returns the flow rate 1 and flow rate 2 for each step (in Multi-
Step Mode).