47
Chemyx, Inc.
Fusion Series User Manual
sets whether the pump is infusing or withdrawing (Fusion 200
only) and the number of steps used in Multi-Step Mode.
X must be an integer or decimal (up to five decimal places)
within the volume limits for the syringe (See read limit parameter
command). A negative sign can be added in front of any value
to indicate a switch to withdrawal mode (Fusion 200 only).
For Basic Mode, only one value is required. Entering only one
value will automatically switch the pump to Basic Mode.
Example
set volume 25.0
To set volumes in Multi-Step Mode, the pump must be
manually switched to Multi-Step Mode on the pump. For each
step, a value must be entered, separated by commas and no
spaces. The number of steps will be automatically adjusted to
the number of values entered (i.e., entering only three values
will change the number of steps to three).
Example
set volume 3,-2,4
Returns the transfer volume, transfer rate, and transfer time (in
Basic Mode).
Returns only the transfer volumes (in Multi-Step Mode).
set time X,X,X,...
Sets the target time for a pump run. This will adjust the transfer
rate so that the run will last for the input time. The target volume
will remain the same. This command also sets the number of
steps used in Multi-Step Mode.
X must be an integer or decimal (up to five decimal places). The
resulting transfer rate must be within the rate limits for the
syringe (See read limit parameter command).
For Basic Mode, only one value[ is required. Entering only one
value will automatically switch the pump to Basic Mode.
Example
set time 1.0
To set target times in Multi-Step Mode, the pump must be
manually switched to Multi-Step Mode on the pump. For each
step, a value must be entered, separated by commas and no