ECO 2 Ultrasound Diagnostic System
Page 118 / 189
2
Power Led lights but LED
no image
The interval time is too short
to restart
Restart after 1 minute
3
LED display character
menu but no scan image
Launch power, gain or STC
control errors
Not connect to probe or the
probe connection is not
correct
Device is in freezing condition
Control launch power ,
gain or STC control
Make sure of right
connection
Exit from freezing
condition
4
Abnormal image
Exam mode errors
Image processing setting
errors
Whether Exam mode is
proper or not
Adjust image processing
setting or set it as default
5
Probe works improperly
1.The plug plugs loosely
2.Internal circuit protects
1.Extract the probe and
reinsert
2.Restart
6
No OB calculation
package menu
Do not select the OB
application before
scanning.
Select the OB application
7
PRINT-key doesn’t work
1.The connected printer is
approved
2.Printer power is not on
3.Printer is not connect well
1.Change the approved
printer
2.Turn on the printer
3.Connect the printer
again