EasyManua.ls Logo

Chison ECO 2 - PW Fast Measurement

Default Icon
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ECO 2 Ultrasound Diagnostic System
Page 84 / 189
Meas. Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
Pressure(CCA)
mmHg
Formula:
Pressure
CCA×CCA/10000
ICA/CCA
Formula:
ICA/CCA
Flow Volume
Diam
Cm
Refer to B distance meas.
TVI
Cm
Time
S
Refer to M time meas.
HR(Single wave)
Bpm
SV
Ml
Formula:
0.785* Diam* Diam*|TVI|
CO
l/min
Formula:
SV*HR(Single wave)/1000
6.25 PW Abdomen Measurement
Refer to PW general measurement.
6.26 PW OB Measurement
Press [Calc] to enter PW measurement in PW OB mode. Or press the corresponding parameter
control key to switch the OB measurement item.
Meas. Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Umb A
Aorta
Descending Aorta
Uterine Artery(Lt)
Uterine Artery(Rt)
Pulmonary Artery
Vs
cm/s
Refer to M speed meas.
Pressure(s)
mmHg
Formula:
PressureVs×Vs/10000
Vd
cm/s
Refer to M speed meas.
Pressure(d)
mmHg
Formula:
PressureVd×Vd/10000
VMean
cm/s
Refer to M speed meas.
Pressure(VMean)
mmHg
Formula:
PressureVMean×VMean/10000
TVI
Cm
SD
Formula:SDVs/Vd

Table of Contents

Related product manuals