Every time before and after each exam, please clean the probe handle and disinfect the
transvaginal and transrectal probes probe using liquid chemical germicides
If the probe is contaminated with body fluids, you should disinfect the probe after cleaning.
Regard any exam waste as potentially infectious and dispose of it accordingly.
CAUTION:
Since the probe is not waterproof, you should disconnect it from the system before cleaning or
disinfecting.
Before and after each exam, please clean the probe handle and disinfect the transvaginal and
transrectal probes using liquid chemical germicides.
Cleaning
You can clean the transvaginal and transrectal probes to remove all coupling gel by wiping with a
soft cloth and rinsing with flowing water. Then wash the probe with mild soap in lukewarm water.
Scrub the probe as needed and use a soft cloth to remove all visible residues from the
transvaginal probe surface. Rinse the probe with enough clean potable water to remove all visible
soap residues, and let the probe air dry.
CAUTION:
Please remove the cover (if any) before cleaning the probe.(The cover like condom is one time
usable).
When cleaning the TV and TR probes, it is important to be sure that all surfaces are thoroughly
cleaned.
Disinfecting
Cidex is the only germicide that has been evaluated for compatibility with the material used to
construct the probes.
To keep the effectiveness of the disinfection solutions, a thoroughly cleaning must be done to the
probe before the disinfecting, make sure no residues remain on the probe.
Disinfecting Procedure:
1. Following all precautions for storage, use and disposal, prepare the germicide solution