ECO 2 Ultrasound Diagnostic System
Page 64 / 189
group of measurement automatically.
6.2.2 Ellipse
Measurement steps:
1. Press the [Calc] button to enter into measurement. Update the [Ellipse] item in the menu or
press the quick measure button [Ellipse]; it will display a segment “+”icon.
2. Move the “+”icon by cursor, Press [ENTER] to fix the point and the cursor can be moved to
form a round.
3. Press [UPDATE] can exchange the activated point and the fixed point.
4. Move the cursor to the end-point of the ellipse, press [ENTER] to fix the axis, at the same time,
the next axis be updated, and can change the size of the axis by the cursor.
5. Now press [UPDATE] can exit to the step 4.
6. After fixing the next axis, can press [ENTER] to complete the measurement.
7. After the measurement, the result will display in the measurement results area.
8. Repeat the steps from 1~6 to start next “ellipse” measurement. Press [DEL] to delete all the
measurements.
NOTE:
Each group of measurement is limited, if the measurement results beyond, it will begin a new
group of measurement automatically.
6.2.3 Trace
Measurement steps:
1. Press the [Calc] button to enter into measurement. Update the [Trace] item in the menu or
press the quick measure button [Trace]; it will display a segment “+”icon.
2. Move the “+”icon by cursor, Press [ENTER] to fix the point and the cursor can be moved to the
next position.
3. Make the cursor tracing along the edge of required area, the traced line can be not closed.
4. Now press [UPDATE] to cancel the tracing.
5. Press [ENTER] again in the endpoint, the start point and end point of trace line will be closed by
a straight line.
6. After the measurement, the result will display in the measurement results area.
7. Repeat the steps from 1~5 to start next “trace” measurement. Press [DEL] to delete all the
measurements.
NOTE:
Each group of measurement is limited, if the measurement results beyond, it will begin a new