EasyManua.ls Logo

Chison ECO 2 - Normal Measurement in M, B;M Mode

Default Icon
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ECO 2 Ultrasound Diagnostic System
Page 79 / 189
Meas. Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
LVIDd
cm
Refer to M distance Meas.
Left Ventricular
Diameter at diastole
LVIDs
cm
Refer to M distance Meas.
Left Ventricular
Diameter at systole
LVPWd
cm
Refer to M distance Meas.
Left Ventricular Posterior
Wall at diastole
IVSs
cm
Refer to M distance Meas.
Interventricular Septal
Thickness at Systole
LVPWs
cm
Refer to M distance Meas.
Left Ventricular Posterior
Wall at Systole
EDV
ml
EDV
7.0/(2.4+LVIDd)×LVIDd3
Left Ventricular Volume
at end diastole
ESV
ml
ESV
7.0/(2.4+LVIDs)×LVIDs3
Left Ventricular Volume
at end Systole
SV
ml
SV|EDV-ESV|
Stroke Volume
SI
SISV/BSA
Stroke Volume,
BSA: Body Surface
Area, it calculated by
input the height and
weight.
EF
%
EFSV/EDV×100
Ejection Fraction
FS
%
FS
(LVIDd-LVIDs)/LVIDd×100
Contraction Fraction
CO
L/min
COSHR/1000
Cardiac Output
CI
CICO/BSA
ECG Index,
BSA: Body Surface
Area, it calculated by
input the height and
weight.
LVMW
LVMW
1.04×[(IVSd+LVIDd+LVPW
d)3-LVIDd3]-13.6
LVMWI
LVMWILVMW/BSA
BSA: Body Surface
Area, it calculated by
input the height and
weight.

Table of Contents

Related product manuals