ECO 2 Ultrasound Diagnostic System
Page 75 / 189
Meas. Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
Ratio
Refer to distance Meas.
Formula: R=D1/D2
D1: First Distance
D2: Second Distance
6.10 B Mode Vessel Measurement
Refer to normal measurement in B mode.
6.11 Urology Measurement
Normally urology measurements are performed in B and 2B mode.
Freeze the required image under Urology examination, then press [Calc] to enter into Urology
measurement status.
Meas. Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
Kidney
Vol.(L/R)
ml
Refer to distance Meas.
Formula: V(π/6)×L×H×W
L: Kidney_L
H: Kidney_H
W: Kidney_w
Bladder Vil.
V(L*W*H)
ml
Refer to distance Meas.
Formula: V(π/6)×L×H×W
L: Bladder _L
H: Bladder _H
W: Bladder _w
Prostate
Volume
ml
Refer to distance Meas.
Formula: V(π/6)×L×H×W
L: Prostate _L
H: Prostate _H
W: Prostate _w
PPSA
ng/ml
Formula: PPSA0.12×V
PSAD
ng/ml
Formula: PSAD=SPSA/V
SPSA: input the SPSA
When create a new
Patient
RVU
Volume
ml
Refer to distance Meas.
Formula: V(π/6)×L×H×W
L: RVU _L
H: RVU _H
W: RVU _w