ECO 2 Ultrasound Diagnostic System
Page 3 / 189
Do not continuously scan the same part of a patient or expose the patient to prolonged
scanning; otherwise it may harm the patient.
When using the system for ultrasound testing, use only qualified ultrasound gel that complies
with system standards.
Do not unplug probe when the system is in active operation. Always go to EXAM screen when
need to remove the probe.
To prevent from arm or neck injury, the operator should not stay at the same position for too
long during patient scanning without taking break.
Do not put liquid on top of the main unit.
NOTE:
*The system has built-in screen saver to avoid the tic mark on the display. It is not recommended
to constantly turn on and off the unit.
*To dispose of this product properly, please call your local service department.
2.2 Electrical Safety
Type of protection against electric shock
Class I Equipment
CLASS I EQUIPMENT in which protection against electric shock does not rely on BASIC
INSULATION only, but includes a protective earth ground. This additional safety precaution
prevents exposed metal parts from becoming LIVE in the event of an insulation failure.
NOTE: The mains supply shall be cut off after disconnecting the power line and the net
power.
Degree of protection against electric shock
Type BF Applied part (for Probes marked with BF symbol)
TYPE BF APPLIED PART providing a specified degree of protection against electric shock, with
particular regard to allowable LEAKAGE CURRENT.