EasyManua.ls Logo

Chison ECO 2 - Page 132

Default Icon
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ECO 2 Ultrasound Diagnostic System
Page 123 / 189
Avoid contacting with solutions or coupling gels containing mineral oil or lanolin.
Avoid temperatures above 60°C. Under no circumstances should the transducer be subjected
to heat sterilization method. Exposure to temperatures above 60º C will cause permanent damage
to the transducer.
Inspect the probe prior to use for damage or degeneration to the housing, strain relief, lens and
seal. Do not use a damaged or defective probe.
Probe Immersion Levels
CAUTION:
These transducers are not designed to withstand heat sterilization methods. Exposure to
temperatures in excess of 60 ºC will cause permanent damage. The transducers are not designed
to be totally submerged in fluid, as permanent damage will result if the entire transducer is
submerged.
Recommended materials
The following specific recommendations are made for the use of Endocavitary ultrasound
transducers. Users should also review the Centers for Disease Control and Prevention document
on sterilization and disinfection of medical devices to be certain that their procedures conform to
the CDC principles for disinfection of patient care equipment.
1. Cleaning Compatibility
Solution
Origin
Qualified Use
Active Ingredients
CIDEZYME
All
Soaking
Enzyme
Epizyme-Rapid(Rapid Multi-Enzyme
cleaner)
All
Soaking
Enzyme

Table of Contents

Related product manuals