EasyManua.ls Logo

Chison ECO 2 - Page 151

Default Icon
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ECO 2 Ultrasound Diagnostic System
Page 142 / 189
intended for transcranial or neonatal cephalic uses.
NOTE 4 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
NOTE 5 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column
related to MI.
NOTE 6 The depths z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply to SCANNING MODES.
Transducer Model: C3-A
Operating Mode: PW
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.60
0.40
0.70
Index component value
0.35
0.40
0.66
0.70
Acoustic
Parameter
s
p
r,α
at Z
MI
Mpa
1.00
P
mW
36.00
36.00
P
1x1
mW
5.17
5.17
Z
s
cm
2.50
Z
b
cm
4.95
Z
MI
cm
4.46
z
pii,α
cm
5.00
f
awf
MHz
2.89
3.90
2.89
Other
Informatio
n
prr
Hz
4386.00
srr
Hz
n
pps
1
I
pa,α
at z
pii,α
W/cm
2
36.31
I
spta,α
at z
pii,α
or
z
sii,α
mW/cm
2
35.20
I
spta
at z
pii
or z
sii
mW/cm
2
58.54
p
r
at z
pii
Mpa
1.60
Operating
Control
Conditons
Focus
cm
9.00
6.00
13.00
Power
%
100
100
100
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS
or TIB.

Table of Contents

Related product manuals