ECO 2 Ultrasound Diagnostic System
Page 9 / 189
PATIENT ENVIRONMENT.
CAUTION:
DO NOT connect any probes or accessories without approval by CHISON within the PATIENT
ENVIRONMENT.
DO NOT touch patient and devices without EN 60601-1 approval to avoid the leakage current
risk within the PATIENT ENVIRONMENT.
Unapproved Devices
CAUTION:
DO NOT use unapproved devices.
If devices are connected without the approval of CHISON, the warranty will be INVALID.
The system can’t be used with HF surgical equipment; otherwise the burns to patient may
occur.
Any device connected to this system must conform to one or more of the requirements listed
below:
IEC standard or equivalent standards appropriate to devices.
The devices shall be connected to PROTECTIVE EARTH (GROUND).
CAUTION:
Unsafe operation or malfunction may result. Use only the accessories, options and supplies
approved or recommended in these instructions for use.
Peripheral used in the patient environment
The system has been verified for overall safety, compatibility and compliance with the following
on-board image recording devices:
B/W video printer: SONY UP-X898MD
The system may also be used safely while connected to devices other than those recommended
above if the devices and their specifications, installation, and interconnection with the system
conform to the requirements of EN60601-1.
Adapter is considered as a part of ME equipment