Page #124 background imageLoading...
Page #124 background image
Programmable DC Electronic Load 63600 Series Operation & Programming Manual
4-48
Local operation:
SAMPLING_TIME:00.100ms
SAMPLING_POINT:3596
TRIG_SOURCE:LOAD_ON
TRIG_POINT:0500
Remote operation:
DIGitizing:SAMPling:TIME 100μs
DIGitizing:SAMPling:POINt 3596
DIGitizing:TRIGger:POINt 500
DIGitizing:TRIGger:SOURce 0
STEP 2: Initializing the function of Digitizing
It is necessary to do initialization before capturing a new waveform. The
initialization action will restart the Digitizing function and capture all Trigger Points
to wait for the Trigger source.
Local operation:
DIGI:INITIATE
Remote operation:
DIGitizing:INITiate
STEP 3: Setting the execution status of Digitizing
IDLE: It indicates the Digitizing is done.
PRE_TRIG: It indicates the Digitizing is processing Trigger Points.
WAIT_TRIG: It indicates the Digitizing is waiting for the trigger signal.
POST_TRIG: It indicates the Digitizing is processing Sampling Points.
Local operation:
It shows <PRE_TRIG…> & <WAIT_TRIG…> under the DIGI:INITIATE.
Remote operation:
DIG:TRIG?
STEP 4: Executing the trigger condition
Refer to “Set the Trigger Source of Digitizing Function.”
STEP 5: Downloading the waveform
(1) Execute DIG:WAV:CAP? Command to send the MODULE waveform to
FRAME. The transmission is done when OK is returned.
(2) Execute DIG:WAV:DATA? V command to download the voltage waveform to
PC from FRAME.
Execute DIG:WAV:DATA? I command to download the current waveform to
PC from FRAME.
To leave out of the Digitizing Function edit page of system configuration, you need to press
and simultaneously to go back to the page of system configuration
When Setting the Configuration is over, to leave out of the page of system configuration, you
need to press and simultaneously to quit the Setting page of system
configuration.
ENTER
ENTER
ADVA
ADVA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63600 Series and is the answer not in the manual?

Chroma 63600 Series Specifications

General IconGeneral
Series63600
Operating Temperature0°C to 40°C
ModesCC, CV
InterfaceGPIB, USB
ProtectionOCP, OTP, OVP
Voltage Resolution1mV
CoolingForced Air

Related product manuals