Local Operation
4. Set the Level of the equivalent series resistance Rs
The setting range is from 30mΩ to 20Ω. There are 15,000 discrete steps in the range.
Press the key to enter into the editing mode. Turn the Rotary knob to change
the display value to 0.15Ω, then press key to confirm.
Rs : 00.15 Ω
5. Review and update the values of the setting parameters
Review the values of the setting parameters by pressing key. Update them by
set new value in their setting pages.
After completion of the data edit, key must be pressed. Otherwise, pressing the
key, new data will not be written into the internal memory, the previous value for
the parameter is kept.
6. Quit from editing mode
Press to quit from editing mode. Then, the VFD display will go back to the
voltage, current, and peak plus/minus voltage display mode.
4.5 Setting Operation Mode of Dynamic Load
4.5.1 Setting the Operation Mode to CC Mode
Dynamic load is only operation in CC mode. Press the key repeatedly until the VFD
displays CC mode.
0 . 0000 mV
0 . 0000 mA
0 . 0000 mW
The sequence of mode selection after pressing key is as follows:
CC CR CV CP CZ goes back to CC