Programmable DC Electronic Load 63600 Series Operation & Programming Manual
5.3.2.21 SYSTEM Subsystem
SYSTem:ERRor?
Type: All Channels
Description: This command queries the error string of the command parser.
Setting Syntax: None
Setting Parameters: None
Query Syntax: SYSTem:ERRor?.
Return Parameters: <ACCRD>, 0,"No Error",
1,Data Format Error",
2,Data Range Error",
3,Command Error",
4,Execution Error",
5,Too Many Errors"
Query Example: SYST:ERR?
SYSTem:REMote
Type: All Channels
Description: This command can only be used under control of USB and Ethernet.
If SYST:REM is programmed, the 63600 will be set in the REMOTE
state, and the front panel of frame will be disabled except the
<LOCAL>key pressed.
Setting Syntax: SYSTem:REMote
Setting Parameters: None
Setting Example: SYST:REM
SYSTem:LOCal
Type: All Channels
Description: This command can only be used under control of USB and Ethernet.
If SYST:LOC is programmed, the 63600 will be set in the LOCAL
state, and the front panel will work.
Setting Syntax: SYSTem:LOCal
Setting Parameters: None
Setting Example: SYST:LOC
M
Type: All Channels
Description: Set the load mode to the ten channels in one frame. The frame will
ignore the setting if the channel does not exist.
Setting Syntax: M<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR1>, 0: do not change, 1: CCL, 2: CCM, 3: CCH, 4: CRL, 5:
CRM, 6: CRH, 7: CVL, 8: CVM, 9: CVH, 10: CPL, 11: CPM, 12:
CPH, 13: CZL, 14: CZM, 15: CZH, 16: CCDL, 17: CCDM, 18: CCDH,
19: CCFSL, 20: CCFSM, 21: CCFSH, 22: TIML, 23: TIMM, 24:
TIMH, 25: SWDL, 26: SWDM, 27: SWDH, 28: OCPL, 29: OCPM, 30:
OCPH, 31: PROG, 34: MPPTL, 35: MPPTM, 36: MPPTH, 37:
UDWL, 38: UDWM, 39: UDWH
Example: M “1,1,2,2,2,2,5,5,0,0”
M ”2,2,2,2,2,2”
AC
Type: All Channels
Description: Set the current level 1(L1) of CC mode to the ten channels in one
frame. The frame will ignore the setting if the channel does not exist.