Remote Operation
Setting Syntax: AC<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=Ampere]
Example: AC “1.0,1,2.5,5.0,10.5,4.5,2.0,2.0,0,0”
AR
Type: All Channels
Description: Set the resistance level 1(L1) of CR mode to the ten channels in one
frame. The frame will ignore the setting if the channel does not exist.
Setting Syntax: AR<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=OHM]
Example: AR “1.0,0.1,0.2,0.5,0.15,0.4,0.2,0.2,0,0”
AV
Type: All Channels
Description: Set the voltage level 1(L1) of CV mode to the ten channels in one
frame. The frame will ignore the setting if the channel does not exist.
Setting Syntax: AV<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=Volt]
Example: AV “5.0,5.5,3.3,5.1,12.0,5.5,5.0,5.2,0,0”
AP
Type: All Channels
Description: Set the power level 1(L1) of CP mode to the ten channels in one
frame. The frame will ignore the setting if the channel does not exist.
Setting Syntax: AP<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=Watt]
Example: AP “50.0,100.0,30,5.1,12.0,5.5,5.0,5.2,0,0”
CCR
Type: All Channels
Description: Set the rising slew rate of CC mode to the ten channels in one frame.
The frame will ignore the setting if the channel does not exist.
Setting Syntax: CCR<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=A/μs]
Example: CCR “1.0,2.5,2.5,10,2.0,5.0,5.0,5.0,0,0”
CCF
Type: All Channels
Description: Set the falling slew rate of CC mode to the ten channels in one
frame. The frame will ignore the setting if the channel does not exist.
Setting Syntax: CCF<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=A/μs]
Example: CCF “1.0,2.5,2.5,10,2.0,5.0,5.0,5.0,0,0”
CPR
Type: All Channels
Description: Set the rising slew rate of CP mode to the ten channels in one frame.
The frame will ignore the setting if the channel does not exist.
Setting Syntax: CPR<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=A/μs]
Example: CPR “1.0,2.5,2.5,10,2.0,5.0,5.0,5.0,0,0”
CPF
Type: All Channels